List of ventilator, CPAP, and BiPAP devices containing PE-PUR foam material in the breathing gas pathway
We have identified additional devices (other than those supplied by Philips Electronics Australia Ltd) that contain PE-PUR foam material for sound abatement purpose in the breathing gas pathway.
We are reviewing all ventilator, continuous positive airway pressure (CPAP), and bi-level positive airway pressure (BiPAP) devices in the Australian Register of Therapeutic Goods (ARTG) to determine whether the devices incorporate harmful sound abatement foam.
This action follows the identification of risks posed by a polyester-based polyurethane (PE-PUR) foam used as soundproofing material in devices supplied by Philips Electronics Australia Ltd.
During the review we identified additional devices (other than those supplied by Philips Electronics Australia Ltd) that contain PE-PUR foam material for sound abatement purpose in the breathing gas pathway.
We requested evidence regarding the long-term safety to ensure these devices are safe for use throughout their expected service life.
Although these additional devices contain PE-PUR foam material, there are different kinds of PE-PUR foam and it may not be the same material grade used in the devices supplied by Philips Electronics Australia Ltd.
We have carefully reviewed each device to understand the composition of the foam used and whether it poses safety risks over time. Each device has been assessed individually to ensure it meets the relevant safety and performance requirements.
Below is the list of other devices (i.e. not supplied by Philips Electronics Australia Ltd) that contain PE-PUR foam material for sound abatement purposes with final outcomes of any regulatory action the TGA may have taken.
We have taken regulatory actions where non-compliances were identified. Currently there are no PE-PUR foam containing devices being supplied in Australia.
Published: Information about Philips' recall action of correcting their affected devices
| ARTG entry | Sponsor | Manufacturer | Device type | Foam type | Models | Regulatory Actions/outcomes |
|---|---|---|---|---|---|---|
| 328181 | PB Medical Pty Ltd | Metran Co Ltd | Home CPAP unit | PE-PUR | METRAN JPAP Travel CPAP Machines | The Sponsor voluntarily cancelled the ARTG entry. PB Medical Pty Ltd has completed recall of the devices already in market. Recall action is Recall; further information can be located by searching the DRAC database for 'METRAN JPAP Travel CPAP Machines' at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx. |
| 370364 | Planet Innovation Pty Ltd | Planet Innovation Pty Ltd | Ventilator, adult | PE-PUR | Juno 200 | The Sponsor voluntarily cancelled the ARTG entry. Planet Innovation has completed recall of the devices already in market. Recall action is Recall; further information can be located by searching the DRAC database for 'Juno 200' at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx. |
| 152125 | Icon Healthcare IP Pty Ltd | F Stephan GmbH | Ventilator, neonatal/paediatric | PE-PUR | EVE NEO | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 254752 | Icon Healthcare IP Pty Ltd | F Stephan GmbH | Transport electric ventilator | PE-PUR | EVE TR | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 260182 | Icon Healthcare IP Pty Ltd | F Stephan GmbH | Ventilator, adult | PE-PUR | EVE IN | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 206250 | Boian Surgical Pty Ltd | Hoffrichter GmbH | Home CPAP unit | PE-PUR | Point 2 Auto CPAP | The device is no longer being supplied and has been cancelled by the Sponsor. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma LAB | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma 25S-C | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma 25S | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma 25ST | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma 30ST | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM100TD -prisma CR | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM110TD -VENT 30-C | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM110TD -VENT 40 | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM110TD -VENT 50 | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | PE-PUR | WM110TD -VENT 50-C | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART Plus | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART Max | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT Plus | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT Max | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM100TD - prisma 20A | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM100TD - prisma 20C | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. As of 2024, this model is no longer supplied in its previous configuration (see [link] for new models supplied). If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNOsoft | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNOsoft 2e | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNObalance | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNObalance e | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART Plus | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SMART Max | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT Plus | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM090TD - prisma SOFT Max | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM100TD - prisma 20A | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | WM100TD - prisma 20C | The manufacturer of this device has undertaken a design change to remove the PE-PUR foam. All new devices are supplied with a different foam. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNOsoft 2 | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNOsoft 2e | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNObalance | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 326866 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | PE-PUR | SOMNObalance e | The device is no longer being supplied. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV56SE | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx. |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV55SE | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx. |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV54AU | Submission is under review. |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV54AU-HH | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV54UK-P | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV54SE-P | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV54SE-HH | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV53SE | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV53SE-HH | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV51SE | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV51D | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV51AU-HH | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV51AU | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 277439 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Home CPAP unit | PE-PUR | DV51SE-HH | The sponsor has provided insufficient evidence of long-term safety to substantiate compliance with the essential principles. A recall has been initiated to remove the devices from the market. Further information can be located by searching the DRAC database for 'RC-2025-RN-01052-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 325776 | Getinge Australia Pty Ltd | Air LiquidMedical Systems SA | Transport electric ventilator | PE-PUR | Monnal T60 Transport Ventilator | The device is no longer being supplied and has been cancelled by the Sponsor. The recall has been initiated. Further information can be located by searching the DRAC database for 'RC-2025-RN-00711-1' under market action at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx |
| 140966 | Draeger Australia Pty Ltd | Draeger AG & Co KGaA | Ventilator, adult | PE-PUR | Carina Hospital Ventilator | Draeger has completed a Product Defect Correction for devices in market. Further information can be located by searching the DRAC database for 'Carina ventilator' at https://apps.tga.gov.au/PROD/DRAC/arn-entry.aspx. |
| 178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 560 Ventilator - 409660 | The Sponsor did not provide sufficient information to substantiate compliance with the Essential Principles. The ARTG entry has been cancelled from the ARTG by the TGA. The recall has been completed for the devices in the market. Further information can be located by searching the Cancellations and Suspensions database for 'ARTG 178669' at https://www.tga.gov.au/resources/cancellations-and-suspensions. |
| 178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 560 Ventilator - 409660-06 | The Sponsor did not provide sufficient information to substantiate compliance with the Essential Principles. The ARTG entry has been cancelled from the ARTG by the TGA. The recall has been completed for the devices in the market. Further information can be located by searching the Cancellations and Suspensions database for 'ARTG 178669' at https://www.tga.gov.au/resources/cancellations-and-suspensions. |
| 178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 520 Ventilator - 4098300 | The Sponsor did not provide sufficient information to substantiate compliance with the Essential Principles. The ARTG entry has been cancelled from the ARTG by the TGA. The recall has been completed for the devices in the market. Further information can be located by searching the Cancellations and Suspensions database for 'ARTG 178669' at https://www.tga.gov.au/resources/cancellations-and-suspensions. |
| 178669 | Medtronics Australia Pty Ltd | Covidien llc | Ventilator, home care | PE-PUR | Puritan Bennett 520 Ventilator - 4098300-06 | The Sponsor did not provide sufficient information to substantiate compliance with the Essential Principles. The ARTG entry has been cancelled from the ARTG by the TGA. The recall has been completed for the devices in the market. Further information can be located by searching the Cancellations and Suspensions database for 'ARTG 178669' at https://www.tga.gov.au/resources/cancellations-and-suspensions. |
| 309191 | AA-Med Pty Ltd | Somnetics International Inc also doing business as Trancend Inc | Home CPAP unit | PE-PUR filter | Transcend 365 miniCPAP Auto | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 309191 | AA-Med Pty Ltd | Somnetics International Inc also doing business as Trancend Inc | Home CPAP unit | PE-PUR filter | Transcend 3 miniCPAP Auto | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 309191 | AA-Med Pty Ltd | Somnetics International Inc also doing business as Trancend Inc | Home CPAP unit | PE-PUR filter | Transcend 3 miniCPAP | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277261 | Zoll Medical Australia Pty Ltd | Zoll Mecial Corporation | Transport electric ventilator | PE-PUR filter | 731 Ventilator EMV+ (800-EMVP) | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277261 | Zoll Medical Australia Pty Ltd | Zoll Mecial Corporation | Transport electric ventilator | PE-PUR filter | 731 Ventilator Zvent (8660-001400, 8660-001401, 8660-001402) | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277261 | Zoll Medical Australia Pty Ltd | Zoll Mecial Corporation | Transport electric ventilator | PE-PUR filter | 731 Ventilator Eagle II (800-EGL2) | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277261 | Zoll Medical Australia Pty Ltd | Zoll Mecial Corporation | Transport electric ventilator | PE-PUR filter | 731 Ventilator AEV (800-AEV1) | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 277261 | Zoll Medical Australia Pty Ltd | Zoll Mecial Corporation | Transport electric ventilator | PE-PUR filter | 754 Ventilator (800-0754) | The Sponsor provided sufficient information to substantiate compliance with the Essential Principles. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
Page history
Data table updated September 2026.
Data table updated February 2026.
Data table updated December 2025
Data table updated September 2025.
Data table updated August 2025.
First published.
Data table updated September 2026.
Data table updated February 2026.
Data table updated December 2025
Data table updated September 2025.
Data table updated August 2025.
First published.