The TGA is reviewing all ventilators, continuous positive airway pressure (CPAP), and bi-level positive airway pressure (BiPAP) devices in the Australian Register of Therapeutic Goods (ARTG) to determine whether the devices incorporate harmful sound abatement foam. This action follows the identification of risks posed by polyester-based polyurethane (PE-PUR) foam used as soundproofing material in devices supplied by Philips Electronics Australia Ltd.
The TGA has identified devices included in the ARTG that contain a different foam material (not PE-PUR foam) for sound abatement purpose in the breathing gas pathway. The TGA has requested evidence regarding the long-term safety to ensure these devices are safe for use throughout their expected service life; some of this information is still under review by the TGA.
For sponsors who have not provided additional evidence to demonstrate the long-term safety of their devices for the intended service life of the devices, the TGA has imposed conditions of inclusion (COI) on the ARTG entries to ensure the timely availability of this evidence to support their devices.
Below is the list of devices that contain other foam material for sound abatement purposes. The TGA will update the list with current status as the review progresses. If hazards resulting in risks to health are identified for these devices, the TGA will undertake appropriate product safety notification or recall action.
Published: Information about Philips' recall action of correcting their affected devices.
| ARTG entry | Sponsor | Manufacturer | Device type | Foam type | Models | Regulatory actions/outcomes |
|---|---|---|---|---|---|---|
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma LAB | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma 25S-C | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma 25S | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma 25ST | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma 30ST | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM100TD -prisma CR | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM110TD -VENT 30-C | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM110TD -VENT 40 | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM110TD -VENT 50 | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 218526 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, bi-level | Polyether polyurethane | WM110TD -VENT 50-C | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SMART | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SOFT | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SMART Plus | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SMART Max | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SOFT Plus | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM090TD - prisma SOFT Max | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM100TD - prisma 20A | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 216631 | Lowenstein Medical Australia Pty Ltd | Lowenstein Medical Technology GmbH Co Kg | Positive airway pressure unit, continuous auto | Polyether polyurethane | WM100TD - prisma 20C | The Sponsor provided sufficient information to support the long term safety of the device. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 165305 | Compumedics Limited | Compumedics Limited | Positive airway pressure unit, continuous auto | Polyether polyurethane | Somnilink SPAP | The device is no longer being supplied and has been cancelled by the Sponsor. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 319236 | CPAP Sales Pty Ltd | Micomme Medical Technology Development Co Ltd | Home CPAP unit | Polyether polyurethane | HealthGear SleepZone C2 (CPAP) | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 319236 | CPAP Sales Pty Ltd | Micomme Medical Technology Development Co Ltd | Home CPAP unit | Polyether polyurethane | HealthGear SleepZone C5 (Auto) | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 187874 | Device Technologies Australia Pty Ltd | Hamilton Medical AG | Transport electric ventilator | Polyether polyurethane | T1 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 187874' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 113999 | Device Technologies Australia Pty Ltd | Hamilton Medical AG | Ventilator, adult | Polyether polyurethane | C1 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 187874' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 113999 | Device Technologies Australia Pty Ltd | Hamilton Medical AG | Ventilator, adult | Polyether polyurethane | MR1 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 187874' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 113999 | Device Technologies Australia Pty Ltd | Hamilton Medical AG | Ventilator, adult | Polyether polyurethane | C3 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 187874' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 113999 | Device Technologies Australia Pty Ltd | Hamilton Medical AG | Ventilator, adult | Polyether polyurethane | C6 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 187874' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 190907 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Positive airway pressure unit, continuous auto | Silicone | DV64E-UK | Sponsor has been requested to provide further information to demonstrate the long term safety of the device. |
| 190907 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Positive airway pressure unit, continuous auto | Silicone | DV64E-G | Sponsor has been requested to provide further information to demonstrate the long term safety of the device. |
| 190907 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Positive airway pressure unit, continuous auto | Silicone | DV64E-SE | Sponsor has been requested to provide further information to demonstrate the long term safety of the device. |
| 190907 | Drive Devilbiss Australia Pty Ltd | Devilbiss Healthcare (USA) | Positive airway pressure unit, continuous auto | Silicone | DV63E-UK | Sponsor has been requested to provide further information to demonstrate the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | RESmart CPAP System - RESmart | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | RESmart Auto CPAP system - RESmart Auto | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Luna @ CPAP System - E-20C-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Noctivance C100 CPAP System (Custom made name) - E-20C-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Luna @ Auto CPAP System - E-20A-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Luna @ Auto CPAP System - E-20AJ-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Noctivance A110 Auto CPAP System (Custom made name) - E-20A-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Noctivance A110 Auto CPAP System (Custom made name) - E-20AJ-H-O | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | CPAP System - G2S C20 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Auto CPAP System - G2S A20 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | CPAP System - G3 C20 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Auto CPAP System - G3 A20 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292407 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home CPAP unit | Polyether polyurethane | Auto CPAP System - M1 Mini | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | RESmart BPAP SYStem -25 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | RESmart BPAP SYStem-25A | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | RESmart BPAP SYStem-25T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | RESmart BPAP SYStem-30T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-20S | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-20A | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-20T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-25S | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-25A | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-25T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System T-30T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System Y-20T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System Y-25T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System Y-30T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System Y-30AT | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System U-20T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System U-25T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System U-30T | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Luna @ BPAP System U-30AT | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B20A | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B25S | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B25A | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B25VT | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B30VT | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 B30SV | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | BPAP System G3 LAB | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Respiratory Insufficiency Ventilator R-90S | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Respiratory Insufficiency Ventilator R-80S | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Respiratory Insufficiency Ventilator R-80B | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 292408 | Emergo Asia Pacific Pty Ltd T/a Emergo Australia | BMC Medical Co Ltd | Home BPAP unit | Polyether polyurethane | Respiratory Insufficiency Ventilator R-80C | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 227979 | Fisher & Paykel Healthcare Pty Ltd | Fisher & Paykel Healthcare Ltd | Home CPAP unit | Polyether urethane | SPSAAA | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 227979' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…. |
| 227979 | Fisher & Paykel Healthcare Pty Ltd | Fisher & Paykel Healthcare Ltd | Home CPAP unit | Polyether urethane | SPSCAA | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 227979' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…. |
| 227979 | Fisher & Paykel Healthcare Pty Ltd | Fisher & Paykel Healthcare Ltd | Home CPAP unit | Polyether urethane | ICONAAA | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 227979' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…. |
| 227979 | Fisher & Paykel Healthcare Pty Ltd | Fisher & Paykel Healthcare Ltd | Home CPAP unit | Polyether urethane | ICONNAA | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 227979' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…. |
| 227979 | Fisher & Paykel Healthcare Pty Ltd | Fisher & Paykel Healthcare Ltd | Home CPAP unit | Polyether urethane | ICONPBA | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data Further information can be located by searching the ARTG database for 'ARTG 227979' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a…. |
| 162468 | Getinge Australia Pty Ltd | Maquet Critical Care AB | Ventilator, adult | Polyether polyurethane | Servo-air | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 337723 | Hospital Products Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV600 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 337723 | Hospital Products Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV800 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 337723 | Hospital Products Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV300 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 223382 | Mindray Medical Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV600 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 223382 | Mindray Medical Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV800 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 223382 | Mindray Medical Australia Pty Ltd | Shenzhen Mindray Bio Medical Electronics Co Ltd | Ventilator, adult | Polyether polyurethane | SV300 | Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 333747 | Rapp Australia Pty Ltd | Beijing Aeonmed Co Ltd | Intensive-care ventilator, general-purpose | Polyether polyurethane | Aeonmed VG70 | The device is no longer being supplied and has been cancelled by the Sponsor. The review has been closed. If the TGA becomes aware of signals or safety issues, the review will be re-opened and the TGA will undertake further investigations. |
| 219429 | Resmed Pty Ltd | Resmed Pty Ltd | Portable ventilator, electric | Silicone | Stellar 100 | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 219429 | Resmed Pty Ltd | Resmed Pty Ltd | Portable ventilator, electric | Silicone | Stellar 150 | A Condition of Inclusion (COI) was imposed on the ARTG entry, requiring the Sponsor to provide additional data to support the long term safety. The Sponsor has subsequently provided additional data and the COI has been removed. |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | S9 AutoSet | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | S9 Elite | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | S9 Escape | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | AirSense 10 AutoSet | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 222485' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | AirSense AutoSet For Her | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data.Further information can be located by searching the ARTG database for 'ARTG 222485' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | AirSense 10 CPAP | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data.Further information can be located by searching the ARTG database for 'ARTG 222485' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 222485 | Resmed Pty Ltd | Resmed Pty Ltd | Home CPAP unit | Polyether polyurethane | AirSense 10 Elite | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 222485' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | AirCurve 10 CS Pacewave | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 236085' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | Lumis 100 VPAP S | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 236085' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | Lumis 150 VPAP ST | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 236085' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | Lumis 150 VPAP ST-A | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 236085' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | S9 VPAP S | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | S9 VPAP ST | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | S9 VPAP ST-A | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | S9 VPAP Adapt | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 236085 | Resmed Pty Ltd | Resmed Pty Ltd | Home BPAP unit | Polyether polyurethane | S9 VPAP Tx | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 304987 | Sefam Medical Pty Ltd | SEFAM | Home CPAP unit | Polyether polyurethane | S.Box (CPAP/Auto-CPAP) | A Condition of Inclusion (COI) was imposed on the ARTG entry requiring the Sponsor to provide additional data to support the long term safety. The COI has been removed as the Sponsor provided sufficient evidence to support the long term safety of the device. |
| 332260 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Ventilator, adult | Polyether, melamine and polyurethane. Polyester air filter | VIVO 50 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 332260' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 332260 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Ventilator, adult | Polyether, melamine and polyurethane. Polyester air filter | VIVO 60 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 332260' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 332260 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Ventilator, adult | Polyether, melamine and polyurethane. Polyester air filter | VIVO 65 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 332260' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 332260 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Ventilator, adult | Polyether, melamine and polyurethane. Polyester air filter | VIVO 30 | This model has been removed from the scope of the review as it has been phased out worldwide and no longer attached to ARTG listing. |
| 332260 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Ventilator, adult | Polyether, melamine and polyurethane. Polyester air filter | VIVO 40 | This model has been removed from the scope of the review as it has been phased out worldwide and no longer attached to ARTG listing. |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Melamine covered with polpropylene. Polyester air inlet filter | VIVO 1 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 317548' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Melamine covered with polpropylene. Polyester air inlet filter | VIVO 2 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 317548' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Melamine covered with polpropylene. Polyester air inlet filter | VIVO 3 | Condition of Inclusion (COI) imposed on the ARTG entry to provide long term safety data. Further information can be located by searching the ARTG database for 'ARTG 317548' at https://www.tga.gov.au/products/australian-register-therapeutic-goods-a… |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Polyurethane and melamine. Polyester air inlet filter | isleep20 | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Polyurethane and melamine. Polyester air inlet filter | isleep20i | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |
| 317548 | The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia | Breas Medical AB | Home CPAP unit | Polyurethane and melamine. Polyester air inlet filter | isleep20+ | The device is no longer being supplied and has been removed from the ARTG entry by the Sponsor. The review has been closed. |