In 2019, we commenced a post-market review of all breast implants included in the Australian Register of Therapeutic Goods (ARTG) in response to the increasing number of cases of breast implant associated anaplastic large cell lymphoma (BIA-ALCL).
The review included laboratory assessment of all implants supplied in Australia and statistical analysis of supply information and cases of BIA-ALCL.
The laboratory report is available on the TGA website.
This webpage details the steps we took to improve the safety of breast implant products in Australia and actions taken as part of this post-market review.
Current status of breast implants previously supplied or being supplied are outlined in the table below.
These include:
- Cancellation (removal) of devices from the ARTG (by the TGA or the Sponsor); or
- Imposition of Conditions of Inclusion (COI) on the devices in the ARTG.
Where a device has been cancelled from the ARTG, this means no new patients will receive these devices and they can no longer be supplied in Australia.
To ensure the highest level of patient safety possible, we have imposed 'conditions' on all breast implants that remain available in Australia.
This means that for a sponsor to continue supplying a device in Australia, they must meet a range of requirements.
These conditions are also imposed on all new breast implant included in the ARTG.
A condition can impose further requirements as seen appropriate to support the safe ongoing use.
The conditions enable us to closely monitor the devices overtime and take action promptly if any issues are identified.
Conditions imposed
- All cases of BIA-ALCL must be reported to the TGA within 10 days of being notified.
- Sponsors are required to provide six monthly reports covering supply data, details of all adverse events, and all complaints, both in Australia and world-wide.
The below conditions have been removed as these conditions have been met:
- The risk of BIA-ALCL must be included in clinicians’ Instructions for Use of all breast implant and tissue expander devices.
- Patient information leaflets must include warnings about the risks of BIA-ALCL
Legacy implant
- Legacy implants are breast implants that women may have had implanted but are no longer available.
- This could be for various reasons; for example, the company may have removed them from the market themselves, or the TGA may have cancelled them for other reasons such as manufacturing concerns.
- Not all legacy implants are listed here, only those which Australian consumers may still have implanted. Breast implants are not a life-time device and are designed to be replaced every 10 to 15 years.
The table below outlines the current status of breast implant devices in Australia.
| Sponsor (Manufacturer) | Device/ARTG number | Product name | Status |
|---|---|---|---|
| Allied Scientific Products Pty Ltd (Nagor Ltd) | 142863 | Nagor Mammary Implants Gel-filled-TEXTURED | Cancelled by the Sponsor |
| 277757 | 9-cell CoGel gel-filled Nagor mammary implant range - textured | Cancelled by the Sponsor | |
| 277758 | Impleo gel-filled Nagor mammary implant range - textured | Cancelled by the Sponsor | |
| 142860 | Gel Filled Mammary Implant | Cancelled by the Sponsor | |
| 277759 | Impleo gel-filled Nagor mammary implant range | Conditions imposed | |
| Emagin Pty Ltd (Groupe Sebbin SAS) | 309613 | Anatomical Breast Implants - Textured - High Cohesive Gel | Cancelled by the Secretary |
| 309614 | Textured, Round, Firm, High Cohesive Gel Mammary Implant | Cancelled by the Sponsor | |
| 309615 | Textured, Round, Soft Classic, Cohesive Gel, Mammary Implant | Cancelled by the Sponsor | |
| 309616 | Microtextured, Round, Firm High Cohesive Gel, Mammary Implants | Cancelled by the Sponsor | |
| 309617 | Microtextured, Round, Classic, Cohesive Gel, Mammary Implants | Cancelled by the Sponsor | |
| 309618 | Smooth, Round, Firm High Cohesive Gel, Mammary Implants | Cancelled by the Sponsor | |
| 309619 | Smooth, Round, Classic, Cohesive Gel, Mammary Implants | Cancelled by the Sponsor | |
| Euro Implants Pty Ltd (Eurosilicone SAS) | 132040 | Cristaline I Aptex/Vertex Paragel Natural Cohesive Gel Implant | Cancelled by the Sponsor |
| 132036 | Paragel Cohesive Gel Implant | Cancelled by the Sponsor | |
| 132037 | Cristaline Paragel Cohesive Gel Implant | Cancelled by the Secretary | |
Euro Implants Pty Ltd (Polytech Health & Aesthetics GmbH) (Formerly JT Medical Pty Ltd) | 171781 | Sublime Line, Smooth, Silicone gel filled Mammary Implants | Conditions imposed |
| 171783 | Sublime Line, Textured, silicone gel filled Mammary Implants | Conditions imposed | |
| 185059 | 4Two Line, Single Lumen, Textured, Silicone gel filled Mammary Implants | Conditions imposed | |
| JT Medical Pty Ltd (Polytech Health & Aesthetics GmbH) | 171782 | Sublime Line, Microthane, Silicone gel filled Mammary Implants | Cancelled by the Secretary |
| 185060 | 4Two Line, Single Lumen, Micro Polyurethane, Silicone gel filled Mammary Implants | Cancelled by the Secretary | |
Life Healthcare Pty Ltd (Establishment Labs SA) (Formerly Spiran Pty Ltd) | 282776 | Sterile Silicone Breast Implants Motiva Implant Matrix SilkSurface PLUS with Qid | Cancelled by the Secretary |
| 282777 | Sterile Silicone Breast Implants Motiva Implant Matrix Ergonomix Round SilkSurface with Qid | Cancelled by the Secretary | |
| 282778 | Sterile Silicone Breast Implants Motiva Implant Matrix SilkSurface PLUS | Conditions imposed | |
| 373251 | Motiva Implants Anatomical Truefixation SmoothSilk Sterile Silicone Breast Implants with Qid - Silicone gel-filled breast implant, smooth-surface | Cancelled by the Secretary | |
| 492425 | Motiva Ergonomix2 SmoothSilk - Silicone gel-filled breast implant, smooth-surface | Conditions imposed | |
| 510445 | Motiva Ergonomix Round SilkSurface/Smoothsilk without Qid - Silicone gel-filled breast implant, smooth-surface | Conditions imposed | |
| Johnson & Johnson Medical Pty Ltd (Mentor / Mentor Medical Systems BV) | 110588 | Siltex Round Cohesive I | Conditions imposed |
| 110589 | Siltex Round Cohesive II | Conditions imposed | |
| 110592 | Siltex Round Becker Cohesive I | Conditions imposed | |
| 119809 | Siltex Contour Becker Cohesive II | Conditions imposed | |
| 130678 | Siltex Contour Gel Breast Implants Cohesive III | Conditions imposed | |
| 226977 | Mentor® CPX™4 Breast Tissue Expander | Conditions imposed | |
| 226982 | Mentor® CPX™4 with Suture Tabs Breast Tissue Expander | Conditions imposed | |
| 110587 | Smooth Round Cohesive I | Conditions imposed | |
| 119646 | Smooth Saline Mammary Prostheses with Diaphragm Valve | Conditions imposed | |
| 319150 | Smooth Round XTRA Gel Breast Implants | Conditions imposed | |
| 326144 | Siltex Round XTRA Gel Breast Implants | Conditions imposed | |
| 393728 | Smooth MemoryGel™ XTRA Breast Implants - Silicone gel-filled breast implant, smooth-surface | Conditions imposed | |
| 393729 | Smooth MemoryGel™ Breast Implants - Silicone gel-filled breast implant, smooth-surface | Conditions imposed | |
Abbvie Pty Ltd (Formerly Allergan Australia Pty Ltd) | 218869 | BRST Round Microcell Textured Responsive gel filled breast implant | Cancelled by the Sponsor |
| 220900 | BRST Round Microcell Textured Cohesive gel filled breast implants | Cancelled by the Sponsor | |
| 171393 | Natrelle Truform1 Gel, Smooth Single Lumen Breast Implants | Cancelled by the Sponsor | |
| 175421 | Natrelle INSPIRA, Truform 1 gel, Smooth, Single Lumen Breast Implants | Cancelled by the Sponsor | |
| 175426 | Natrelle INSPIRA Truform 2 gel, Smooth, Single Lumen Breast Implants | Cancelled by the Sponsor | |
| 220696 | BRST Round Smooth Responsive gel filled breast implants | Cancelled by the Sponsor | |
| Removed from ARTG 2019 | Biocell Natrelle range macro-textured implants | Legacy implant | |
| Device Technologies Australia Pty Ltd (Silimed Industria de Implantes LTDA) | Removed from ARTG 2016 | Mammary Implant-Silicone Gel-Polyurethane Foam Coated | Legacy implant
|
| Mammary Implant-Silicone Gel-Smooth Surface | Legacy implant | ||
| Mammary Implant - Silicone Gel - Textured Surface | Legacy implant | ||
Emergo Asia Pacific Pty Ltd T/a Emergo Australia (Airxpanders Inc) | Removed from ARTG 2019 | Aeroform Patient Controlled Tissue Expander | Legacy implant |
| AeroForm™ Tissue Expander System Dosage Controller Kit - Remotely-controlled tissue expander remote control | Legacy implant | ||
| Medical Vision Aust Plastic & Cosmetic Pty Ltd (Cereplas) | Removed from ARTG 2014 | CEREFORM Silicone Gel-Filled Breast Implant (Smooth Texture) | Legacy implant |
| CEREFORM Silicone Gel-Filled Breast Implant (Intermediate Texture) | Legacy implant | ||
| CEREFORM OPTIMA Silicone Gel-Filled Breast Implant | Legacy implant | ||
| Medical Vision Australia (Poly Implant Protheses (Pip) SA) | Removed from ARTG 2010 | Poly Implant Prosthese PIP | Legacy implant |
| Surgitek | Removed from ARTG 2002 | Polyurethane-coated breast implant | Legacy implant |