Purpose
The purpose of this document is to provide manufacturers and sponsors with guidance on the Therapeutic Goods Administration’s (TGA) expectations concerning clinical performance requirements (i.e., clinical sensitivity and specificity) and risk mitigations for in vitro diagnostic medical devices (IVDs) intended to be used as self-tests for Chlamydia trachomatis (chlamydia), Neisseria gonorrhoeae (gonorrhoea) and Treponema pallidum (syphilis).
This document identifies key risks that must be mitigated and identifies conditions that may be imposed on these self-test kits if they are to be included in the Australian Register of Therapeutic Goods (ARTG).
Additional risks and mitigation strategies, including conditions of inclusion, may apply to individual devices on a case-by-case basis.
Other aspects of demonstrating safety and performance, such as analytical performance studies or stability studies are not addressed in this document.